Product Category
Tygon® ND 100-80 Tubing
Tygon® ND 100-80 tubing has met the requirements of Biological Test for Plastics, Class VI, as described in USP <88> (2017; fluid path), making it well-suited for many medical applications. The relatively high durometer of this thermoplastic formulation imparts a higher resistance to pressure and enables manufacturing at small diameters (< 0.03”/1 mm OD) and wall thickness (as compared to lower durometer alternatives). The physical properties of Tygon ND 100-80 make our microbore tubing an ideal choice for drainage system, diagnostic equipment, and pressurized fluid transfer.
Tygon® ND 100-80 Tubing Product Details
Features & Benefits
- Fluid path has met the test requirements of Biological Test for Plastics, Class VI, as described in USP <88> (2017)
- Clear for easy visualization of the fluid flow
- Microbore tubing available (< 0.03”/1 mm OD)
- Pressure resistance
- Physical properties are not significantly impacted by the following sterilization methods: autoclave (steam at 135°C during 30 min), EtO, and gamma radiation
- Tygon ND 100-80 is designed and manufactured without the intentional addition of animal-derived materials or DEHP
Typical Applications
- Drug delivery (e.g., intravenous or intra-arterial infusion)
- Cannulas or protective sheaths
- External drainage system (e.g., cerebral spinal fluid drainage)
- Diagnostic equipment (e.g., reagent manufacturing, waste transfer)
- Pressurized fluid transfer
Physical Properties
| Property | ASTM Method | Value or Rating |
|---|---|---|
| Durometer Hardness (Shore A), 15 sec. | D2240 | 80 |
| Tensile Strength, psi (MPa) | D412 | 2,625 (18.1) |
| Ultimate Elongation, % | D412 | 350 |
| Tear Resistance, lb-f/inch (kN/m) | D1004 | 275 (48.2) |
| Specific Gravity | D792 | 1.22 |
| Water Absorption, % 24 hrs at 73°F (23°C) | D570 | 0.10 |
| Compression Set Constant Deflection, % @ 158°F (70°C) for 22 hours |
D395 Method B | 59 |
| Maximum Recommended Operating Temperature °F (°C) | - | 185 |
| Brittleness by Impact Temperature, °F (°C) | D746 | -31 (-35) |
| Tensile Modulus @ 100% Elongation, psi (MPa) | D412 | 1,450 (10.0) |
*Unless otherwise noted, all tests were conducted at room temperature (73°F).
**Values shown were determined on 0.075" thick extruded strips, 0.075" thick molded ASTM plaques, or molded ASTM durometer buttons.
Standard Dimensions
| Part Number | ID (inch) | OD (inch) | WT (inch) | ID (mm) | OD (mm) | WT (mm) | Length (feet) | Length(m) |
|---|---|---|---|---|---|---|---|---|
| AAD04091 | 0.01 | 0.03 | 0.01 | 0.25 | 0.76 | 0.255 | 500 | 152 |
| AAD04103 | 0.02 | 0.06 | 0.02 | 0.51 | 1.52 | 0.505 | 500 | 152 |
| AAD04119 | 0.03 | 0.09 | 0.03 | 0.76 | 2.29 | 0.765 | 500 | 152 |
| AAD04127 | 0.04 | 0.07 | 0.015 | 1.02 | 1.78 | 0.38 | 500 | 152 |
| AAD04133 | 0.05 | 0.09 | 0.02 | 1.27 | 2.29 | 0.51 | 500 | 152 |