Product Category
Tygon® MPF-300 Tubing
Tygon® MPF-300 tubing is a clear, high-performance polymer not formulated with PVC or DEHP, however, trace levels of DEHP are not controlled. This flexible tubing is designed to replace PVC in medical devices in, which DEHP plasticizers and leachables are not desired. Tygon MPF-300 has excellent bonding strength and is ideally suited for fluid and oxygen administration applications. Tygon MPF-300 is developed specifically for medical applications and tested to various elements of ISO 10993 including hemocompatibility, implantation, irritation and skin sensitization, systemic toxicity, and cytotoxicity.
Increasing concerns in Europe and North America regarding health hazards and environmentally friendly disposal of DEHP plasticizers led to the development of the MPF product line. Tygon MPF-300 is designed and manufactured without the intentional addition of DEHP.
Tygon MPF-300 Product Details
Features & Benefits
- PVC replacement - Tubing not made with plasticizers
- Excellent chemical resistance
- Cleanroom manufacutring available
Physical properties are not significantly impacted by the following sterilization methods:
- Gas - EtO (Ethylene Oxide)
- Radiation (25 kGy/2.5 Mrad)
Physical Properties
| Property | ASTM Method | Value or Rating |
|---|---|---|
| Durometer Hardness (Shore A), 15 sec | D2240-02 | 83 |
| Tensile Strength, psi (MPa) (at break) | D412-98 | 3,595 (41.1) |
| Ultimate Elongation, % | D412-98 | 553 |
| Tear Resistance, lb-f/inch (kN/m) | D624-91, Die C | 509 (83.1) |
| Compression Set Constant Deflection, % @ 158°F (70°C) for 22 hours |
D395-02, Method B | 29 |
| Brittle Temperature, °F (°C) | D746-98 | < -112 (-80) |
| Specific Gravity | D792-02 | 1.11 |
| Water Absorption, % 24 hrs at 73°F (23°C) | D570-98 | 1.13 |
| Maximum Recommended Operating Temperature °F (°C) | n/a | 180 (82) |
| Tensile Modulus @ 100% Elongation, psi (MPa) | D412-98 | 1.546 |
Characteristics
Tygon MPF-300 has met the following test requirements:
- Plastic Class VI, as described in USP <88> (2016)
- Hemocompatibility, as described in ISO 10993-4 (2022)
- Implantation, as described in ISO 10993-6 (2022)
- Skin Sensitization, as described in ISO 10993-10 (2022)
- Acute Systemic Toxicity, as described in ISO 10993-11 (2022)
- Irritation, as described in ISO 10993-23 (2022)
*Product recommendations are based on a combination of industry knowledge, material science expertise, and/or material testing data. Contact Saint-Gobain Medical for further tubing recommendation information.